Ir directamente a la navegación principal Ir directamente a la búsqueda Ir directamente al contenido principal

Design and construction of a magnetic resonance compatible multi-injector gas jet delivery system

  • David Megias-Alguacil
  • , Thierry Keller*
  • , Kai Lutz
  • , Ashley P. Barlow
  • , Dominik A. Ettlin
  • *Autor correspondiente de este trabajo

Producción científica: Contribución a una revistaArtículorevisión exhaustiva

5 Citas (Scopus)

Resumen

We present the design, construction, and performance of a novel multi-injector gas jet delivery capable of operating in a magnetic resonance imaging environment. This apparatus is computer controlled and built with two separate pneumatic circuits enabling gas jet applications at variable sites through four independently activated injectors. Gas jet delivery is fully controllable in terms of pressure, flow rate, gas temperature, application time, and duration of interstimulus interval. We characterized these parameters, considering effects such as pressure drop by flow transport, transient effects, and delays in activation. The system offers new possibilities for use in various biomedical contexts such as, e.g., quantitative sensory testing or dental hypersensitivity assessment.

Idioma originalInglés
Número de artículo014301
PublicaciónReview of Scientific Instruments
Volumen79
N.º1
DOI
EstadoPublicada - 2008
Publicado de forma externa

Huella

Profundice en los temas de investigación de 'Design and construction of a magnetic resonance compatible multi-injector gas jet delivery system'. En conjunto forman una huella única.

Citar esto