Real-time chest-wall-motion tracking by a single optical fibre grating: A prospective method for ventilator triggering

  • M. D. Ivanović
  • , J. Petrovic
  • , A. Savić
  • , G. Gligorić
  • , M. Miletić
  • , M. Vukčević
  • , B. Bojović
  • , Lj Hadžievski
  • , T. Allsop
  • , D. J. Webb
  • University of Belgrade
  • Aston University

Producción científica: Contribución a una revistaArtículorevisión exhaustiva

2 Citas (Scopus)

Resumen

Objective: The ventilators involved in non-invasive mechanical ventilation commonly provide ventilator support via a facemask. The interface of the mask with a patient promotes air leaks that cause errors in the feedback information provided by a pneumatic sensor and hence patient-ventilator asynchrony with multiple negative consequences. Our objective is to test the possibility of using chest-wall motion measured by an optical fibre-grating sensor as a more accurate non-invasive ventilator triggering mechanism. Approach: The basic premise of our approach is that the measurement accuracy can be improved by using a triggering signal that precedes pneumatic triggering in the neuro-ventilatory coupling sequence. We propose a technique that uses the measurement of chest-wall curvature by a long-period fibre-grating sensor. The sensor was applied externally to the rib-cage and interrogated in the lateral (edge) filtering scheme. The study was performed on 34 healthy volunteers. Statistical data analysis of the time lag between the fibre-grating sensor and the reference pneumotachograph was preceded by the removal of the unwanted heartbeat signal by wavelet transform processing. Main results: The results show a consistent fibre-grating signal advance with respect to the standard pneumatic signal by (230 ± 100) ms in both the inspiratory and expiratory phases. We further show that heart activity removal yields a tremendous improvement in sensor accuracy by reducing it from 60 ml to 0.3 ml. Significance: The results indicate that the proposed measurement technique may lead to a more reliable triggering decision. Its imperviousness to air leaks, non-invasiveness, low-cost and ease of implementation offer good prospects for applications in both clinical and homecare ventilation.

Idioma originalInglés
Número de artículo045009
PublicaciónPhysiological Measurement
Volumen39
N.º4
DOI
EstadoPublicada - 26 abr 2018
Publicado de forma externa

Huella

Profundice en los temas de investigación de 'Real-time chest-wall-motion tracking by a single optical fibre grating: A prospective method for ventilator triggering'. En conjunto forman una huella única.

Citar esto