Resumen
According to the Medical Devices Directive, both the preparation for clinical trials and marketing of implants require that a risk analysis is performed. This paper presents a risk analysis for a dental implant in the framework of the risk management process carried out for the preparation of a multi-centre clinical trial, where likely hazards, failure modes and their severities, probabilities and detectabilities are assessed, together with a review of the related scientific literature. The clinical study aimed to evaluate a new ion implantation-based implant surface designed for the promotion of more extensive and faster osseointegration.
Idioma original | Inglés |
---|---|
Páginas (desde-hasta) | 1201-1204 |
Número de páginas | 4 |
Publicación | Medical Engineering and Physics |
Volumen | 30 |
N.º | 9 |
DOI | |
Estado | Publicada - nov 2008 |