Ir directamente a la navegación principal Ir directamente a la búsqueda Ir directamente al contenido principal

Uncertainty analysis of the new Spanish National Pressure Standard

  • A. Herrero*
  • , M. C. Matilla
  • , N. Cereceda
  • , S. Ruiz
  • , R. Bueno
  • *Autor correspondiente de este trabajo

Producción científica: Contribución a una revistaArtículo de revisiónrevisión exhaustiva

2 Citas (Scopus)

Resumen

This paper presents the project for developing a new interferometer-based mano-barometer. This system will constitute the National Primary Pressure Standard maintained at CEM. This device is based on the ancient Torricelli's experience and applies modern techniques in order to obtain a very accurate instrument. This system will allow the calibration up to 300 kPa with uncertainties about 5 · 10-6. Laser interferometry is applied and special attention is given to the accurate determination of fluid density. Hence, this paper describes the technology used in the project. Its influence on the measurement is being studied prior to the manufacturing and assembling phases.

Idioma originalInglés
Páginas (desde-hasta)467-470
Número de páginas4
PublicaciónCIRP Annals - Manufacturing Technology
Volumen51
N.º1
DOI
EstadoPublicada - 2002
Publicado de forma externa

ODS de las Naciones Unidas

Este resultado contribuye a los siguientes Objetivos de Desarrollo Sostenible

  1. ODS 9: Industria, innovación e infraestructura
    ODS 9: Industria, innovación e infraestructura

Huella

Profundice en los temas de investigación de 'Uncertainty analysis of the new Spanish National Pressure Standard'. En conjunto forman una huella única.

Citar esto